Electronic speckle pattern shearing interferometer with a photopolymer holographic grating

Research output: Contribution to journalArticlepeer-review

Abstract

A photopolymer holographic grating is used to produce the two sheared images in an electronic speckle pattern shearing interferometer. A ground glass screen following the grating eliminates unwanted diffraction orders and removes the requirement for the CCD camera to resolve the diffraction grating's pitch. The sheared images on the ground glass are further imaged onto the CCD camera. The fringe pattern contrast was estimated to be above 90%. A validation of the system was done by comparing the theoretical phase difference distribution with the experimental data from the three-point bending test.

Original languageEnglish
Pages (from-to)2439-2442
Number of pages4
JournalApplied Optics
Volume43
Issue number12
DOIs
Publication statusPublished - 20 Apr 2004

Fingerprint

Dive into the research topics of 'Electronic speckle pattern shearing interferometer with a photopolymer holographic grating'. Together they form a unique fingerprint.

Cite this